Cuvette Holder for 10-cm CuvettesThe CUV-UV-10 Cuvette Holder for 10-cm path
length cuvettes couples via SMA-terminated optical fibers to Ocean Optics
high-sensitivity miniature fiber optic
spectrometers and light sources to create small-footprint
spectrophotometric systems for absorbance/transmission measurements of
aqueous solutions and gases. This compact cuvette holder is optimized for
UV-VIS-NIR (~200 nm-2 µm) applications.
CUV-UV-10:
$635
The CUV-UV-10 consists of these key
components:
-
Two adjustable 5-mm diameter f/2 quartz
collimating lenses
-
Cell compartment that accepts 10-cm
cuvettes
-
Built-in filter slot
-
Manual light block for dark measurements
-
Internal channels designed to accept a
constant-temperature water source (for heating and cooling of the base
and cuvette, by means of convection)
-
SMA 905 terminations for coupling to
fiber optics.
Each cuvette holder assembly comes with a
black cover, which can be used to eliminate ambient light.
Using the 10-cm cuvette holder is simple:
Attach one SMA-terminated optical fiber to one of the assembly's
collimating lenses and another SMA-terminated optical fiber to the
assembly's other collimating lens. One of these fibers -- the illumination
fiber -- is coupled to a fiber optic light source. The other fiber -- the
read fiber -- is coupled to the spectrometer. In some instances, you may
wish to adjust the collimating lenses for precise positioning of the fiber
at the lens focal point.
"Z" Dimension
Cuvettes and cuvette holders offered by Ocean
Optics have a "Z" dimension of 15 mm. Only the
qpod has a Z dimension of 8.5 mm.
User's Tip
The
collimating lenses common to each cuvette holder can be unscrewed from the
holder assembly and used separately in any fixture for applications that
require a fiber for free-beam coupling, such as on-line transmission or
reflection.
Specifications
|
Pathlength: |
10 cm |
|
Collimating lenses (UV-VIS-NIR): |
»
Quartz lenses optimized
from 200 nm-2 µm
»
5 mm diameter
»
f/2 |
|
Collimating lens termination: |
SMA 905 |
|
Filter slot: |
Accepts filters up to 1/4" (6
mm) in thickness |
|
Base material: |
Aluminum |
|
Connections in base: |
Water inlet/outlet |
|
Water input fittings: |
1/8" NPT pipe thread |
|
Recommended optical fibers*: |
200 µm illumination fiber;
50 µm read fiber |
* There is no single
combination of optical fibers that satisfies the requirements of every
application. As a rule, however, it is best to use a large-diameter
(>50 µm) illumination fiber to get maximum light throughput and a
small-diameter (<50 µm) read fiber to achieve the best optical
resolution.
|